Ứng dụng sản phẩm
- The LTS 670 leak test system can be used to test hollow objects for leaks due to overpressure or vacuum.
Thông số kỹ thuật
- Power supply: U = 230 V, 50 Hz
- Compressed air: clean pressurized air, max. 7 bar
- More powerful power is available upon request.
Adjustment time:
- Filling time: 0.0 ... 3,000,0 seconds
- stable. time: 0.0 ... 3,000,0 seconds
- Test time: 0.0 ... 3,000.0 seconds
- Empty time: 0.0 ... 3,000,0 seconds
Available pressure ranges:
- -1 ... 0 Bar
- 0 ... 1 Bar
- 0 ... 4 Bar
- 0 ... 6 Bar
- 0 ... 10 Bar
Measuring range:
- Relative pressure test: like 1/20000 . resolution pressure range
- differential pressure test: ±20 mbar, resolution 0.2 Pa
Designs:
- IP40 19“, 3HE 450mm x 140mm x 326mm (W x H x D) desktop case, or ½19”, 6HE 235mm x 266mm x 326mm (W x H x D) with separate slots for power and compressed air parts
- Provides connection, interface and main switch on the rear panel
- Serial interface 9600 bps ,1 start bit ,8 data bits (ASCII characters),1 stop no parity check
- Options available
- two external selectable test connections
- Start-up test via external switch
- Mandatory supply voltage
- Tools to support testing and calibration
- Special protocol for serial interface